Tabula: Strep A AntigenRapid Test vs. PCR Test
Methodus | Culture | Summa Proventuum | ||
Strep A Antigen Celeri Test | Proventus | Positivum | Negative | |
Positivum | 102 | 7 | 109 | |
Negative | 6 | 377 | 383 | |
Summa Proventuum | 108 | 384 | 492 |
Sensitivum relativum: 94.4% (88.3%-97.9%)*
Relativum: 98.2% (96.3%-99.3%)*
Cura: 97.4% (95.5%-98.6%)*
* 95% Fiducia Intervalla
Strep A Antigen Test Rapid celeri chromatographica immunoassay pro qualitative detectio Strep A antigenis ab speciminibus swab faucibus auxilio in diagnosi Group A Infectio Streptococcalis.
1.Recognoverunt ut summus technicus in Sinis, plures applicationes pro diplomatibus et programmatibus programmatis probatae sunt.
2.Deliver bona ut ordo petitionem
3.ISO13485, CE Praeparate varia documenta shipping
4.Reply customers quaestiones in XXIV horas